Density Calculation
Slurry Material
C-14
DensityPRO User Manual
Thermo Scientific
–
Measured Inputs
I: Radiation Intensity – Counts per second from the detector. The radiation
intensity shall be the filtered counts.
Istd: Standardization Counts – The average count rate per second from the
detector during standardization. The detector will calculate the average count
rate during the standardization cycle.
Ical1: Calibration point 1 counts – The average count rate per second from
the detector during calibration of the first point. The detector will calculate
the average count rate during the calibration cycle or during the
standardization cycle when standardized on a full pipe.
Ical2: Calibration point 2 counts – The average count rate per second from
the detector during calibration of the second point. The detector will calculate
the average count rate during the calibration cycle.
–
Calculated values
L: Pipe ID (cm) – Inside diameter of the pipe converted from user selected
units to centimeters.
Dcal1: Calibration point 1 Density – Density for calibration point 1 (Dc1u)
converted to grams per cubic centimeter.
Dcal2: Calibration point 2 Density – Density for calibration point 2 (Dc2u)
converted to grams per cubic centimeter.
D: Density of process material – Calculated using
Additional
Measurements
In addition to Measurement #1, three additional measurements, Measurements #2 –
#4 are available to display additional parameters measured or calculated by the gauge.
These additional measurements are also capable of displaying the values from the
analog inputs.
–
–
Solids content/volume (Reference
–
Carrier content/vol (Reference
–
Solids/carrier (Reference
–
Percent by weight solids (Reference
–
Percent by weight carrier (Reference
–
Percent by volume solids (Reference
–
Percent by volume carrier (Reference
–
Bulk Mass Flow rate – The flow rate of the slurry calculated using the density of
the slurry.
Содержание DensityPRO NAI Series
Страница 2: ......
Страница 4: ......
Страница 6: ...This page intentionally left blank ...
Страница 8: ...This page intentionally left blank ...
Страница 14: ...xiv DensityPRO User Manual Thermo Scientific This page intentionally left blank ...
Страница 22: ...A xxii DensityPRO User Manual Thermo Scientific Quick Setup This page intentionally left blank ...
Страница 124: ...Operation Application Screens 4 72 DensityPRO User Manual Thermo Scientific Figure 4 39 High and Low Alarms ...
Страница 134: ......
Страница 158: ......
Страница 195: ......