4-2
DB-070 E
RadEye AB100
Pi/Ff 04.05.2021
Finder mode:
In the Finder mode, the sound frequency depends on the pulse rate
of the detector. The more detector pulses above the count rate that
is present at the moment of the activation of the finder mode are
captured by the detector, the higher the tone. Unlike the single
pulse indication as described in section 4.1, this mode is controlled
by the microprocessor and is thus slightly delayed upon rapid
changes of the count rate.
The finder mode must be enabled in the menu.
It is activated and deactivated by two times pressing the right (
►
)
or top button:
4.2
Alarm indication
Each time the first alarm threshold is exceeded, the alarm devices
beeper, LED and the vibrator become active, if they are enabled.
Alarm 1 αβ threshold:
LED slowly blinking, two frequency alarm tone and vibrator eve-
ry second
Alarm 2 αβ threshold:
LED quick blinking, continuous alarm tone and vibrator every
second
Indication of activated
finder mode