Operation
45
In extreme cases an unsymmetrical
loading of the rotor could release the
imbalance identification.
Imbalance does not only result in noisy
operation but also in a premature wear
of the drive.
Inserting the centrifuge tubes
The rotor has to be loaded symmetrically. When
loading the rotor only partially, you have to ensure
that opposite bore holes always receive tubes of
equal weight (when centrifuging a single sample,
place a centrifuge tube e.g. filled with water).
After placing the tubes, close the rotor lid.
Fixed-angle rotors:
Proper loading
Improper loading
Содержание Cryofuge 5500i GMP
Страница 16: ...Description of the Cryofuge 12 Notes ...
Страница 20: ...Before the use 16 Notes ...
Страница 70: ...Maintenance and care 66 ...
Страница 99: ...Appendix 95 ...
Страница 107: ...Appendix 103 ...
Страница 113: ......