Solenoid Switch and
Indicator
The Power switch (mains disconnect)is located at right rear of the storage
system cabinet. When power is applied to the controller, the red LED
indicator lights.
The Cryo controller automatically controls the liquid nitrogen level
between the low and high level sensors. Liquid nitrogen enters the storage
tank through a remotely-operated, normally closed, electric solenoid valve.
The solenoid valve is mounted inside of cabinet with access panel
provided.
Manual Fill button activates the solenoid valve when the liquid level is
greater than one inch below the high level sensor and will fill until the
high level sensor is submerged in liquid nitrogen. If a higher level is
desired, the Manual Fill button must be held down until that level is
reached.
The yellow LED solenoid indicator lights whenever the fill solenoid valve
activates.
7000740r7 Cryo Series 2-5
Thermo Scientific
Section 2
Operation
Power Switch and
Power Indicator