![Thermo Scientific CC58114PBC Скачать руководство пользователя страница 47](http://html1.mh-extra.com/html/thermo-scientific/cc58114pbc/cc58114pbc_installation-and-operation-manual_1104600047.webp)
Thermo Scientific
8-1
8
Excess Temperature Option
The Excess Temperature Option, when installed, provides an additional,
independent temperature control system to help protect products from excess
temperatures.
Read this section carefully before using this option.
Control Display
When the Excess Temperature Controller Mode UT150L is first turned on, it
displays only the excess temperature setpoint in the bottom display. Press and
release the SET/ENT button to show the duration time of the last excess
temperature incident. (
See Section “Exceeded Temperature Duration Timer”)
Press and release the SET/ENT button once again to show the peak temperature
measured for the last excess temperature incident.
Pressing and releasing the SET/ENT button once again shows the current
temperature measured by the controller in the top display; this may differ slightly
from the main temperature controller.
The value shown in the bottom display is the current excess temperature setpoint.
Pressing and releasing the SET/ENT button again cycles back to the first display
of the SET/OUT in the lower display.
Excess Temperature Option Features
Exceeded Temperature Duration Timer
The Exceeded Temperature Duration Timer measures the time that the setpoint is
exceeded (and power to the heaters was interrupted) until the hysteresis value is
reached as the chamber temperature cools. This time indicates when the chamber
temperature exceeded the Excess Temperature setpoint.
Peak Exceeded Temperature
The Peak Exceeded Temperature is the highest temperature measured by the
Excess Temperature Controller.
Содержание CC58114PBC
Страница 1: ......
Страница 6: ...List of Figures ii Thermo Scientific ...
Страница 12: ...Introduction 2 2 Thermo Scientific ...
Страница 14: ...Unpacking 3 2 Thermo Scientific ...
Страница 20: ...Installation Thermocouple Wiring Installation 4 6 Thermo Scientific ...
Страница 22: ...Initial Start up 5 2 Heratherm Thermo Scientific ...
Страница 46: ...Communication Option Addresses for Multiple Controllers 7 6 Thermo Scientific ...
Страница 50: ...Excess Temperature Option Excess Temperature Controller Operation 8 4 Thermo Scientific ...
Страница 57: ...Replacement Parts Thermo Scientific Heratherm 11 3 Figure 5 Wiring Diagram ...
Страница 58: ...Replacement Parts 11 4 Heratherm Thermo Scientific ...
Страница 59: ...Thermo Scientific 12 1 12 Error Codes This chapter is not applicable ...
Страница 60: ...Error Codes 12 2 Thermo Scientific ...
Страница 62: ...WEEE Compliance 13 2 Thermo Scientific ...
Страница 63: ...Thermo Scientific 14 1 14 14 Spare Parts and Accessories This chapter is not applicable ...
Страница 64: ...Spare Parts and Accessories 14 2 Thermo Scientific ...
Страница 66: ...Device Log 15 2 Thermo Scientific ...