Maintenance
Thermocouple Replacement
9-8
Thermo Scientific
2. Note polarity and wire location. Disconnect the positive and negative
thermocouple lead wire extension leads from the terminal screws on the
thermocouple head.
3. Lift the thermocouple vertically out of its hole in the top insulation.
4. Replace the thermocouple and connect the new wires. Red is always negative.
(If the extension leads are black and white, white is negative),
5. Replace the furnace top access panel and replace the screws.
Figure 7. Thermocouple Replacement