Maintenance
Thermocouple Replacement
Thermo Scientific
9-7
Figure 6.Door Insulation Replacement
Thermocouple Replacement
All thermocouples are subject to aging and deterioration. This condition is
indicated by a gradual drop in millivolt output for any temperature and by furnace
operation at a higher temperature than the controller indicates. For critical
processes, periodically check the furnace chamber temperature with a separate
thermocouple and instrument to determine the amount of error.
The appropriate frequency of thermocouple replacement depends upon frequency
of operation and operating temperatures. For additional information, refer to Table.
“UT 150 Controller Operation”
To replace the thermocouple, complete the following steps (refer to
Figure
1. Remove the two top access panel screws and remove the top access panel.
The maul disconnect power does not need to be turned off to
replace the thermocouple
Содержание BF51634BC-1
Страница 1: ......
Страница 6: ...List of Figures ii Thermo Scientific...
Страница 20: ...Safety Notes Standards and Directives 1 14 Thermo Scientific...
Страница 48: ...Communication Option Auto Tuning the Controller 7 10 Thermo Scientific...
Страница 52: ...Excess Temperature Option B Models Excess Temperature Controller Operation 8 4 Thermo Scientific...
Страница 64: ...Troubleshooting 10 4 Thermo Scientific...
Страница 65: ......
Страница 66: ......
Страница 69: ...Thermo Scientific 12 1 12 Error Codes This chapter is not applicable...
Страница 70: ...Error Codes 12 2 Thermo Scientific...
Страница 72: ...WEEE Compliance 13 2 Thermo Scientific...
Страница 73: ...Thermo Scientific 14 1 14 14 Spare Parts and Accessories This chapter is not applicable...
Страница 74: ...Spare Parts and Accessories 14 2 Thermo Scientific...
Страница 76: ...Device Log 15 2 Thermo Scientific...