Installing the Basic System
Installing the Internal Battery
Thermo Fisher Scientific
AutoPILOT PRO Flow Computer User Guide 3-11
With the internal battery option, the instrument is powered by a 12 V,
28 AH rechargeable lead acid battery. The battery assembly consists of the
battery (customer supplied) and the battery bracket kit.
Warning
Ensure power is off and the area is non-hazardous before making
any connections.
Warning
Use wires suitable for at least 90°C.
Warning
Installation must be carried out in accordance with local site
requirements and regulations.
Warning
When used in hazardous locations, only the Power Sonic
PS-12280 battery may be installed.
Note
The internal battery option cannot be used with the 212A modem
option.
Note
The internal battery option is not available for the AutoPILOT PRO
XL.
1. Open the flow computer enclosure. Lift the main board and mounting
bracket to expose the chassis bracket.
2. Refer to the
general assembly drawing
for internal battery assembly
orientation and mounting location in the enclosure.
3. Refer to Figure 3–10 for the following:
a. Place one star lock washer over the mounting holes in the chassis
bracket, and install the standoffs.
b. Fit the battery between the standoffs. Align the holes on the battery
retainer with the standoffs, and secure the retainer with the screws
and lock washers.