Guidelines for handling loaded chips
To ensure the proper function of your loaded chips, handle them according to the
following guidelines:
• Always wear powder-free gloves.
IMPORTANT!
Oils from your hands can contaminate the chips and interfere
with imaging.
• Handle the loaded chips by gently gripping the sides of each chip case, being
careful to not apply pressure to the chip within it.
• Do not touch the window of the chip lid. If you accidentally touch the window,
clean it using a laboratory wipe that has been sprayed with isopropanol, then dry
by wiping in one direction using a clean low-lint wipe.
Note:
Debris or excess sample on the lid can interfere with thermal uniformity
during thermal cycling, or fluoresce during imaging.
• Thermal cycle chips within 2 hours after loading them.
443MP
J
443MP
J
443MPJ
Chapter 4
Thermal cycle the chips
Guidelines for handling loaded chips
4
QuantStudio
™
3D Digital PCR System User Guide
35