Configure ramp rates
The ramp rate is how quickly the temperature changes from one stage or step to another stage or step.
The maximum ramp rate for the instrument is 6.5°C/second.
1.
In the
Method
tab, tap
Edit
4
Advanced
4
Ramp rates
.
Note:
Use or to navigate through the method if it is not all displayed.
2.
Tap a ramp rate, then edit the value.
3.
Tap
Done
.
Add or remove a pause
A pause allows you to open the drawer to add additional reagents to the plate.
1.
In the
Method
tab, tap
Edit
4
Advanced
4
Add pause
.
Note:
Use or to navigate through the method if it is not all displayed.
2.
Tap
(Pause)
.
3.
Enter a temperature and the cycle.
A pause in the thermal cycling protocol is denoted with a . The cycle of the pause and the
temperature of the pause are also denoted.
4.
Tap
(Pause)
to remove the pause.
5.
Tap
Done
.
Options to edit the well details of a plate file
The following files can be edited:
• A system template (see
A system template that is edited must be saved as a plate file. A system template cannot be
overwritten.
• A saved plate file (see “Create and run a plate from a system template or a saved plate file”
• The plate file associated with the last run (see “Repeat the last run”
Chapter 4
Edit a plate before starting a run
Options to edit the well details of a plate file