Operating Instructions 

Alpha RES 1000 

 

 

17

7.4.5

 

Maximum Pulse Length (tPL) or Maximum Frequency (FPF) 

Note: If the controller type “oFF” or “L.Ct” is set, the parameters listed in 7.4.4 and 7.4.5 are blanked out.  
This mode lets you set the maximum pulse length or the maximum frequency at which the relay will operate.  

1)

 

Follow directions in 7.4.1 to enter Controller mode. 

 

2)

 

Press the ENTER key. Scroll until the lower display shows “t.PL” or 
“F.PF”.  

10.0 

t.PL

HOLD  SETUP 

-

 

In PLC (pulse length) mode: The lower display shows “t.PL” to indicate 

pulse length. The upper display shows your current pulse length. You 
can select any value from 0.5 to 20 seconds.  

 

-

 

In PFC (pulse frequency) mode: The lower display shows F.PF to 

indicate pulse frequency. The upper display shows your current 
maximum pulse rate. You can select any value from 60 to 120 pulses 
per minute. When the measured value exceeds the Proportional Band  
in 7.4.4, the controller will pulse the relay at this rate. 

 

60 

F.PF

 

HOLD  SETUP 

3)

 

Press the 

Δ

 and 

 keys to choose the period duration or maximum 

frequency, depending on your mode. 

4)

 

Press the ENTER key to confirm your selection and to return to 
Advanced Set-up mode, or return to Measurement mode by pressing the 

Δ

 and 

 keys simultaneously (escape). 

 

7.5

 

Measurement Range sub-function 

In this sub-function, appropriate range is selected with the appropriate cell constant.  
 
7.5.1

 

Entering the Measuring Range sub-function 

1)

 

Enter Advanced Set-up mode. Push ENTER key and scroll to Advanced Set-up security code “22”. 
Push ENTER key. 

2)

 

Press 

Δ

 or 

 keys to scroll until upper display shows “rng”. 

 

7.5.2

 

Selecting Measuring Range sub-function 

rng

 

 

 

HOLD  SETUP 

1)

 

  Follow directions in 7.5.1 to enter Controller mode and press ENTER 

key

2)

 

Press 

Δ

 or 

 keys to select correct range. (Please refer to 7.5.3 for full 

list of measurement ranges). Note the upper display shows maximum 
measurement range, while lower display shows Cell constant. Lower left-
corner of LCD displays the number corresponding to the respective range. 

3)

 

Press ENTER key to confirm. 

20.00 

0.01

 

HOLD  SETUP 

m

Ω

 

4)

 

Proceed to 7.5.4 to set current output or return to Measurement mode by 

pressing 

Δ

 and 

 keys simultaneously (escape). 

Содержание Alpha RES 1000

Страница 1: ...Alpha RES 1000 Controller Transmitter Resistivity...

Страница 2: ...k ROSS Resolution SAOB SMART AVERAGING Smart CheK SMART STABILITY Stacked Star Navigator 21 Stat Face The Enhanced Lab ThermaSense Triode TRIUMpH Unbreakable pH Universal Access are trademarks of Ther...

Страница 3: ...d applications of the Alpha RES 1000 Resistivity controller transmitter If you have doubts in the use of the instrument please do not hesitate to contact the nearest Thermo Scientific Authorised Distr...

Страница 4: ...D SET UP MODE 10 7 1 TEMPERATURE COEFFICIENT SUB FUNCTION 11 7 1 1 Selecting Pure water or Linear Temperature Coefficient 11 7 2 EMPERATURE CALIBRATION ATC MODE ONLY 12 T 7 2 1 Setting manual temperat...

Страница 5: ...20 7 6 4 Selecting steady or pulse contact for the alarm relay if the relay 3 is set to Alarm 20 7 6 5 Wash Contact if the relay 3 is set to Wash 20 7 6 6 Input Line Resistance Adjust 21 7 6 7 Reverti...

Страница 6: ...C 0 to 1999 second time delay adjustment on all relays minimise false alarms Separately adjustable high and low set point hysteresis dead bands prevent chattering of relays around the set points Selec...

Страница 7: ...process assembly a final control element such as pump or valve and a chart recorder Power Mains 220 110 VAC roller Alpha RES 1000 Cont Flow Assembly to Dosing Pumps Process Assembly with Electrode Ch...

Страница 8: ...Diagram Signal Output put Hold In Alarm Relay 2 Relay 1 Power Mains Res mA PE S 14 15 17 15 16 ror conditions ty PE AC N L 1 2 5 4 3 6 7 8 9 10 11 12 13 indicated contact positions are for currentles...

Страница 9: ...nection 8 Relay 2 relay common 17 0 4 20 mA for ve connection 9 Relay 2 relay working position NO Connections for the 5 way screw terminals 18 Pt1000 Pt100 lead 1 terminal red 19 Pt1000 Pt100 sense le...

Страница 10: ...een RELAY A and RELAY B in MANUAL relay operation mode 4 1 2 Display The LCD display features two numerical displays that show status messages and measured values for easy quick reference The display...

Страница 11: ...OLD Set up parameters can be viewed or changed by entering a security code See Section 5 2 for security code information 4 2 1 How to view operating parameters without access to change them 000 S Cd a...

Страница 12: ...selection of parameters Controller offers three levels of password protection 1 for viewing all SETUP Configuration without the facility to make changes 2 for direct access to calibration function an...

Страница 13: ...to return to Measurement mode 5 1 2 Checking set points To read current set point values without changing them Press RELAY Selection Rel A Rel B key Upper display shows set point for Relay A lower di...

Страница 14: ...ent mode press and keys simultaneously escape NOTE If to view not change set up parameters push ENTER key when security code reads 000 5 2 2 1 Clearing Advanced Setup security code from display Having...

Страница 15: ...libration standard value should be 4 00 m to 19 99 m 4 Once the reading has stabilized use the or keys to adjust the measured value to that of the standard solution Press the ENTER key to accept the v...

Страница 16: ...y code 22 Push ENTER key Controller displays tC Press ENTER key again Controller displays Pur in upper display and tC in lower display Default is Pur 2 Press or keys to select between Pur and Lin temp...

Страница 17: ...he Measurement mode by pressing the and keys escape simultaneously 7 2 1 Setting manual temperature compensation Note This parameter is unavailable if ATC is selected HOLD SETUP oFF For manual tempera...

Страница 18: ...SETUP HOLD SETUP HOLD SETUP HOLD SETUP HOLD SETUP HOLD SETUP HOLD SETUP HOLD SETUP 4 Press the ENTER key to confirm your selection 5 Proceed to 7 3 3 or return to Measurement mode by pressing the and...

Страница 19: ...etely independent of each other 4 Press ENTER key to confirm your selection 5 Proceed to 7 3 5 or return to Measurement mode by pressing the and keys simultaneously escape NOTE Please refer to Appendi...

Страница 20: ...easurement mode by pressing the and keys simultaneously escape 7 4 Controller Cntr sub function You can set the controller s parameters in this sub function 7 4 1 Entering the Controller sub function...

Страница 21: ...r display shows rEL and the upper display shows the current selection de energised dEEN or energised EN 3 Press the or keys to choose de energised or energised relay state 4 Press the ENTER key to con...

Страница 22: ...ess the and keys to choose the period duration or maximum frequency depending on your mode 4 Press the ENTER key to confirm your selection and to return to Advanced Set up mode or return to Measuremen...

Страница 23: ...ay shows the output type 0 20 or 4 20 and the lower display shows out 3 Press the or keys to select your output type 0 20 or 4 20 mA 4 Press ENTER key to confirm your selection 5 Proceed to 7 5 6 or r...

Страница 24: ...Enter Advanced Set up mode Push ENTER key and scroll to Advanced Set up security code 22 Push ENTER key 2 Press or keys to scroll until upper display shows ConF Press ENTER key to enter configuration...

Страница 25: ...ther alarm contact will operate as a steady contact or a fleeting single pulse contact Pulse contact closing time is 1 second Stdy AL C HOLD SETUP 1 Follow directions in 7 6 1 to enter Configuration m...

Страница 26: ...ode no HOLD SETUP 2 Press the ENTER key The upper display shows no or YES the lower display shows deF default 3 Press the or keys to select no or YES WARNING Press the ENTER key resets all settings an...

Страница 27: ...pper display until it reads 22 3 Press ENTER key Manual indicator by the RELAY CONTROL key lights up Note Pressing ENTER key at a value other than 22 will cause controller to revert to Measurement mod...

Страница 28: ...A Alarm Wash Functions Function switchable Latching pulse Pickup delay 0 to 1999 sec Switching voltage max 250 VAC Switching current max 3A Switching power max 600 VA Electrical Data and Connections P...

Страница 29: ...buse Out of warranty products will be repaired on a charge basis 10 1 Packaging The instrument is packaged in a corrugated box with a warranty card instruction manual and the following accessories 17...

Страница 30: ...voltage 110 VAC JP 6 Selects the temperature sensor for Pt1000 Pt100 Fuse Note that there is a fuse slow blow 100mA internal to the controller Before opening the unit ENSURE that the power cable is p...

Страница 31: ...9 mS cm 0 1 1 M cm Potable Water Maximum 1 5 mS cm 0 7 M cm Brackish Water 1 80 mS cm 0 01 1 M cm Industrial Process Water 7 140 mS cm rarely stated Ocean Water 53 mS cm rarely stated Appendix 3 Simp...

Страница 32: ...Operating Instructions Alpha RES 1000 27 Appendix 4 Pure Water Curve Resistivity of Pure Water 0 10 20 30 40 50 60 70 80 90 100 0 5 10 15 20 25 30 35 40 45 50 55 60 65 70 75 80 85 90 95 100 Temp C...

Страница 33: ......

Страница 34: ...Fax 978 232 6031 Europe P O Box 254 3860 AG Nijkerk Wallerstraat 125K 3862 CN Nijkerk Netherlands Tel 31 033 2463887 Fax 31 033 2460832 Asia Pacific Blk 55 Ayer Rajah Crescent 04 16 24 Singapore 13994...

Отзывы: