21
7.3.3
Choosing High or Low set points
Select a “lo” to activate relay when the pH/ORP value undershoots the set point;
select a “Hi” to activate relay when the value overshoots the set point. Using
both SP1 and SP2, the following configuration for the set points can be selected:
lo/lo, lo/hi, hi/lo, or hi/hi.
1)
Follow directions in 7.3.1
to enter Control Relay mode.
2)
Press ENTER key
until upper display shows Lo or Hi and lower display
shows SP1 (SP2).
3)
Press
Δ
or
∇
keys
to select low (lo) or high (hi) set point for SP1 (SP2).
4)
Press ENTER key
to confirm.
5)
Proceed to 7.3.4 step 3
, or return Measurement mode by pressing
Δ
and
∇
keys simultaneously (escape).