10.
Setting the proportional range: press the
▲
or
▼
key
to set the
proportional range (setting range: 10 to 500%).
Press the ENT key
to
confirm your setting.
11.
Setting the pulse frequency: press the
▲
or
▼
key
to set the pulse
frequency (setting range: 60 to 120 pulses per minute
). Press the ENT
key
to confirm your setting.
12.
Setting the integral action time: press the
▲
or
▼
key
to set the
integral action time (setting range: 0.0 to 999.9 minutes).
Press the ENT
key
to confirm your setting.
Note:
If the integral action time is set to ‘0.0’, then the unit functions as a
‘P’ Controller.
Continue with Setup mode procedures, or return to Measurement mode by
pressing the
▲
and
▼
keys simultaneously (escape).
6.6
Current Output 1 sub-function
In this sub-function you set the current output range of the transmitter for pH
or mV values.
2
2
3
3
4
4
5
5
SETUP
HOLD
P
H
SETUP
HOLD
P
H
SETUP
HOLD
P
H
SETUP
HOLD
P
H
SETUP
HOLD
P
H
SETUP
HOLD
SETUP
HOLD
P
H
SETUP
HOLD
1
cur.1
SETUP
HOLD
ENT
ENT
ENT
ENT
ENT
ENT
ENT
ENT
1. Select the “CUR.1” sub-function, then
press the ENT key
.
2.
Selecting the output type:
press the
▲
or
▼
key
to select the desired
output type: 0-20 mA or 4-20 mA.
Press the ENT key
to confirm your
selection.
3.
Setting the pH or ORP value at which the transmitter output will be 4
mA (or 0 mA):
press the
▲
or
▼
key
to set the pH or ORP value to be
equivalent to 4 mA or 0 mA, respectively.
Press the ENT key
to confirm
your setting.
4.
Setting the pH or ORP value at which the transmitter output will be 20
mA: press the
▲
or
▼
key
to set the pH or ORP value to be equivalent
to
20 mA.
Press the ENT key
to confirm your setting.
5.
Selecting out of range current: Press the
▲
or
▼
key
to switch “CU22”
on or off. If “CU22” is on, the output signal will jump to 22 mA when the
measured pH and ORP value is below or above the values set in points 3
& 4 above.
Press the ENT key
to confirm your setting.
Note:
the “CU22” setting (on/off) of current output 1 applies to current
output 2 as well.
Continue with Setup mode procedures, or return to Measurement mode by
pressing the
▲
and
▼
keys simultaneously (escape).