Instruction Manual
Alpha-
COND 550
24
5.6 Viewing Electrode Properties
Each time you calibrate the conductivity sensor, the monitor re-calculates the
calibration factor of the sensor and shown in the LCD at the end of calibration. The
setup mode allows you to view the last saved calibration factor and cell constant at
any time.
SETUP
HOLD
2
SETUP
ENT
1
ENT
HOLD
SETUP
HOLD
ENT
1
1
From measurement mode
press ENT key
to enter setup mode as
described in section 5.1. The LCD shows the first screen of setup mode
(
rAng
).
Press
▲
or
▼
key
to
select viewing electrode properties screen
(
CdAt
).
Press ENT key
to view electrode properties (
CdAt
).
2
Cell constant:
The lower display shows ’
CEL
’ and the upper display
shows the last configured cell constant.
Press ENT key.
3
Calibration factor:
The lower display shows ‘FCt’ and the upper display
shows the calibration factor based on the last conductivity calibration.
NOTE:
The upper display shows ‘
- - -
‘ if the monitor has not been
calibrated yet or the monitor has been reset to factory defaults.
Press ENT key
to quit from electrode properties screen. The monitor
reverts to
CdAt
screen.
Press
▲
or
▼
key
to access other setup screens or
press
▲
and
▼
key
simultaneously
(escape) to return to measurement mode.