ADVANCE D SETUP FUNCTIONS
The advanced setup mode lets you customize your transmitter’s preferences and
defaults. This transmitter features different sub groups that organize all setup
parameters. The sub-groups are:
6.1
Range and Zooming Selection Setting
Your COND 500 transmitter provides six selections of conductivity measurement
ranges to suit your process application needs. Zooming selection is available where
both the 4 mA and 20 mA output current can be assigned to specific conductivity
values for a more refined output.
SETUP
HOLD
SETUP
HOLD
19.99
1
Six Range selection
μ
S
From the measurement mode,
1. Press the ENT key to enter the setup mode.
Transmitter displays ‘
SEt rng
’ indicating the
range setup mode.
2. Press the ENT key again. The display will show
the last set conductivity measurement range.
3. Use
the
▲
or
▼
keys to scroll through the six
ranges and press the ENT key to select your
desired range. The transmitter will now switch to
the zooming selection and display the low zoom
value setup.
NOTE: Range 6 selection page will only appear in the
setup mode when the cell constant is selected as K =
1.0 or K = 10. Range 1 will only appear when cell
constant is selected as K =0.1 or K =1.0