Controller/Transmitter

uS / mS / ºC

Содержание Alpha COND 500

Страница 1: ...Alpha COND 500 Controller Transmitter uS mS C...

Страница 2: ...k ROSS Resolution SAOB SMART AVERAGING Smart CheK SMART STABILITY Stacked Star Navigator 21 Stat Face The Enhanced Lab ThermaSense Triode TRIUMpH Unbreakable pH Universal Access are trademarks of Ther...

Страница 3: ...ransmitter as possible If there are doubts in the use of the transmitter please do not hesitate to contact the nearest Authorized Distributor Thermo Scientific will not accept any responsibility for d...

Страница 4: ...RATURE CALIBRATION 11 5 3 CONDUCTIVITY CALIBRATION 12 6 ADVANCED SETUP FUNCTIONS 13 6 1 RANGE AND ZOOMING SELECTION SETTING 13 6 2 TEMPERATURE COMPENSATION SETTING 15 6 3 HOLD CURRENT SETTING 18 6 4 O...

Страница 5: ...ia the 2 wire power supply loop This transmitter has many user friendly features all of which are completely accessible through the water resistant membrane keypad Your transmitter includes an instruc...

Страница 6: ...re 1 Connection Guide All wiring is done on 2 detachable connectors 1 9 pin connector located on SL1 position for Conductivity electrode and temperature sensor 2 3 pin connector located on SL2 positio...

Страница 7: ...500 2 2 Connecting the Electrode and Temperature Sensor SL1 Position 2 2 1 To connect the Conductivity electrode The Alpha COND 500 transmitter accepts any industrial electrodes for conductivity meas...

Страница 8: ...rt the Pt100 return wire and the shield wire to pin 7 of SL1 connector Tighten the screw 2 wire temperature sensor 1 Take a small length of wire and short pins 5 and 6 of SL1 connector Insert Pt 100 s...

Страница 9: ...3 INSTALLATION 3 1 Mechanical Dimensions 2 wire Conductivity Transmitter Alpha COND 500 3 2 Wall Mount...

Страница 10: ...cut out of 92 0 mm by 92 0 mm 2 Remove back cover of transmitter and slide it through panel cut out Panel side 3 Atta side ch lugs to either of transmitter 4 Thread rods through lugs until transmitter...

Страница 11: ...ar Upper Secondary D Lower Disp y Display Display isplay lay 1 SETUP Setup mode indicator 5 Temperature Coefficient indicator 9 ATC Automatic Temp Compensation 13 Calibration standard indicator 2 MEAS...

Страница 12: ...CAL to enter Conductivity Calibration mode To scroll through the various submenus To increment decrement values or toggle between options in the SETUP CAL modes When pressed together serves as escape...

Страница 13: ...he cell constant is selected to be K 1 0 or K 10 whereas range 1 0 00 to 19 99uS cm will only appear when cell constant is selected to be K 0 1 or K 1 0 Other wise the selection will be limited to fiv...

Страница 14: ...ach time you calibrate your meter NOTE Your transmitter is factory set to a temperature coefficient of 2 1 per C For most applications this will provide good results The factory default value for norm...

Страница 15: ...ll be erased once a new cell constant is selected It is important to ensure that temperature calibration is carried out first prior to conductivity calibration since temperature readings affect the ac...

Страница 16: ...the CAL key to enter calibration mode The upper display will show the measured reading based on the last calibration setting whereas the lower display will show the measured default reading of the sta...

Страница 17: ...values for a more refined output SETUP HOLD SETUP HOLD 1 9 9 9 1 Six Range selection S From the measurement mode 1 Press the ENT key to enter the setup mode Transmitter displays SEt rng indicating th...

Страница 18: ...onfirm The transmitter now displays the high zoom value setup SETUP HOLD 0 00 L SETUP HOLD 1 9 9 9 Hi S S 5 Use the or keys again to set the conductivity high zoom value for the 20 mA current and pres...

Страница 19: ...he ENT key to enter the temperature compensation mode Display shows AtC On or AtC Off Use S or T keys to select AtC On and then press the ENT key 3 Dip ATC sensor into a solution of known temperature...

Страница 20: ...n press the ENT key 3 The primary display will show temperature default is 25 0 and the secondary display shows P C This is the page for setting the temperature of your process or sample by using an a...

Страница 21: ...SETUP HOLD SETUP HOLD SETUP HOLD C ATC SE HO TUP LD SE HO TUP LD SE HO TUP LD Temperature Compensation Setting Chart Alpha COND 500 Transmitter...

Страница 22: ...ent output will be equivalent to last measured value 6 4 Out of Range Current Setting The transmitter can be set to give a fix current output of 3 8 mA when the conductivity readings exceed the zoom s...

Страница 23: ...to enter the setup mode Use the or keys to scroll till LCD displays SEt in the upper display and tPr in the lower display 2 Press the ENT key to view the temperature coefficient setup page tC O 3 Pre...

Страница 24: ...nce Setting Line resistance of an electrode cable constitutes to error in measurement of high conductivity values as in range 6 This error can be compensated by inputting the electrode cable line resi...

Страница 25: ...ration point for the selected measuring range If there is no calibration done in the selected range the display will show 6 9 Viewing the Electrode Data The COND 500 transmitter lets you check the ele...

Страница 26: ...easurement mode 1 Press the ENT key to enter the setup mode Use the or keys to scroll till LCD displays rSt in the upper display 2 Press the ENT key to enter the reset page Use the or keys to toggle b...

Страница 27: ...A or the last current output value HLd OFF SET org Selecting Out of Range current output to be constant 3 8 mA or 4mA for under range and 20 mA for over range org OFF SET tPr Setting Temperature Coeff...

Страница 28: ...terference a Ensure shield wire is properly connected to pin 7 Slow response a Dirty Oily probe a Clean probe Blinking ATC a No temperature sensor connection during ATC mode a Ensure temperature sensi...

Страница 29: ...sation Auto Pt 100 from 0 to 100 C Manual from 0 to 100 C Temperature Coefficient Factor Setting 0 to 10 Normalization Temperature Setting 25 0 o C or 20 0 o C Number of calibration points 1 Output 4...

Страница 30: ...Material SS316 and 25ft cable open ended EC CS10 0 01S Conductivity 2 Cell type probe up to 20 S Cell constant K 0 01 with integrated Pt 100 Material Titanium and 25ft cable open ended EC CS10 0 01T...

Страница 31: ...ument shall not apply to defects resulting from Improper or inadequate maintenance by customer Unauthorized modification or misuse Operation outside of the environment specifications of the products 1...

Страница 32: ...Fax 978 232 6031 Europe P O Box 254 3860 AG Nijkerk Wallerstraat 125K 3862 CN Nijkerk Netherlands Tel 31 033 2463887 Fax 31 033 2460832 Asia Pacific Blk 55 Ayer Rajah Crescent 04 16 24 Singapore 13994...