Operating Instructions
Thermo Scientific Alpha COND 1000
21
7.5
Measurement Range sub-function
In this sub-function, the appropriate range is
selected with the appropriate cell constant.
7.5.1 Selecting
the
Measuring Range sub-
function
1) Enter Advanced Set-up mode. Push the
ENTER
key and scroll to Advanced Set-up security code
“22”. Push the
ENTER
key again.
2) Press
the
▲
or
▼
keys to scroll until the upper
display shows “rng”.
3) Press
the
ENTER
key. The upper display shows
the maximum measurement range, while the lower
display shows the Cell constant. In the lower left-corner of the LCD, the number
corresponding to the respective range is displayed.
4) Press
the
▲
or
▼
keys to select the correct range. (Please refer to Section
12.2 Appendix 2 for the full list of measurement ranges).
5) Proceed to 7.6
to set current output or return to Measurement mode by
pressing the
▲
and
▼
keys simultaneously (escape).
HOLD
SETUP
4
200.0
0.1
μ
S
HOLD
SETUP