5
Maintenance
Preparing a New Syringe
Thermo Scientific
AI 3000/AS 3000 Sampling System - Operating Manual
79
To perform a manual cleaning cycle proceed as follows:
1. Remove the syringe as described in the operating sequence
2. Completely take out the syringe plunger and clean it gently using a soft cloth or a paper
tissue wetted with high cleaning power solvent (acetone is recommended).
3. Reintroduce the syringe plunger and gently move it repeatedly along its whole travel
drawing an organic solvent (e.g. hexane).
4. Reinstall the syringe as described in the operating sequence
Preparing a New Syringe
This operating procedure is always advisable before using a new syringe with low capacity
(like a 5 or 0.5
μ
L) or a new 10
μ
L syringe with Teflon tip plunger. The following procedure
will allow to obtain the proper lubrication of the plunger.
To prepare a new syringe proceed as follows:
1. Fill a 4 mL vial with fresh solvent (hexane is recommended) and place it in the D position
of the washing tray of the AI 3000/AS 3000 II.
2. From your PC, enter the Data System and then access the AI 3000/AS 3000 II method
page. Select the following parameters:
• Pre-injection Solvent: D
• Pre-injection Cycles: 15
3. Enter the
File
menu and click on
Send Method to Sampler
.
4. Enter the
Command
menu and click on
Start Clean Cycle
.