UltiMate 3000 Series:
AFC-3000 Automated Fraction Collector
Page
60
Operating Instructions
Problem
Probable Cause
Remedial Action
The AFC-3000 does
not work correctly
when controlled by
Chromeleon.
The fraction collector is not
correctly configured in
Chromeleon.
Check the configuration for the correct
COM port selection
(
→
(Cont'd)
The USB port on the computer is
not ready for operation.
Check the USB port on the computer.
6.3 Messages in the Chromeleon Audit Trail
When the fraction collector is operated with Chromeleon, messages are displayed in the
Chromeleon Audit Trail.
The following table lists fraction collector-related messages along with appropriate remedial
actions. In addition to the messages in the table, other messages may appear. In this case, note
the exact wording of the message and contact Thermo Fisher Scientific Service for Dionex
HPLC Products if you are unable to eliminate the problem.
Message
Remedial action
Can't open communication channel
Serial Device Driver\COM2.
Check the configuration for the correct COM port selection
(
→
Command for function Connect
failed.
The AFC-3000 cannot connect to Chromeleon. See
Communication Problems (
→
page 59).
Command for function Rack
failed.
Verify that you have configured the racks in the appropriate
order, and that you have not configured more racks than can be
installed (
→
page 49).
Command to change property
TubePosition failed.
You have entered a tube position that exceeds the number of
positions on the installed racks. Verify that you have defined a
valid rack type. Enter a valid tube position.
Device is not remote.
The AFC-3000 is not connected to Chromeleon. Execute the
Connect command.
Error P0040: Value out of range.
Verify that you have entered values that are within the allowed
range.
Internal error: A required
parameter was missing.
Verify that you have defined a legal rack type for all used rack
positions, and
NotUsed
for unused rack positions (
→
page 49).
Invalid enumerated value 3.
Verify that you have entered valid rack types only (
→
page 49).
Parameter TubePosition value
exceeded its allowed range. Set to
closest valid value 384.
You have entered a tube position that exceeds the number of 384.
Enter a valid tube position. This error appears only with the use
of the
96Pos
rack type.
Содержание AFC-3000
Страница 2: ...UltiMate 3000 Series AFC 3000 Automated Fraction Collector Operating Instructions ...
Страница 4: ...UltiMate 3000 Series AFC 3000 Automated Fraction Collector Page II Operating Instructions ...
Страница 18: ...UltiMate 3000 Series AFC 3000 Automated Fraction Collector Page 12 Operating Instructions ...
Страница 28: ...UltiMate 3000 Series AFC 3000 Automated Fraction Collector Page 22 Operating Instructions ...
Страница 48: ...UltiMate 3000 Series AFC 3000 Automated Fraction Collector Page 42 Operating Instructions ...
Страница 76: ...UltiMate 3000 Series AFC 3000 Automated Fraction Collector Page 70 Operating Instructions ...
Страница 78: ...UltiMate 3000 Series AFC 3000 Automated Fraction Collector Page 72 Operating Instructions ...
Страница 82: ...UltiMate 3000 Series AFC 3000 Automated Fraction Collector Page 76 Operating Instructions ...
Страница 86: ...UltiMate 3000 Series AFC 3000 Automated Fraction Collector Page 80 Operating Instructions Sample run ends here ...