UltiMate 3000 Series:
AFC-3000 Automated Fraction Collector
Page
32
Operating Instructions
Fraction rack
Maximum number of racks/bottles
Rack layout
96Pos
Four well plates (96 positions) can be installed
on the well plate adapter. Well plates must not
be mixed with other rack types.
10Pos_WPS,
22Pos_WPS,
40Pos_WPS
Four WPS-3000 well-plate sampler racks can
be installed on the rack adapter. Any four
racks may be used together at one time.
Bottle
Six 250 mL Schott bottles can be placed
directly on the universal drip tray. You must
always install all six bottles. It is not possible
to mix bottles with other rack types.
Foxy rack positioning pins
When Foxy Jr./R1 racks are used, Thermo Fisher Scientific recommends installing two
positioning pins (part no. 6702.9006) to prevent incorrect installation and orientation of a
Foxy rack. Place the pins in the holes provided in the universal drip tray. With the pins
installed, a Foxy rack cannot be installed incorrectly on the universal drip tray.
Fig. 10: Positioning pins for Foxy racks
Positioning pins
Содержание AFC-3000
Страница 2: ...UltiMate 3000 Series AFC 3000 Automated Fraction Collector Operating Instructions ...
Страница 4: ...UltiMate 3000 Series AFC 3000 Automated Fraction Collector Page II Operating Instructions ...
Страница 18: ...UltiMate 3000 Series AFC 3000 Automated Fraction Collector Page 12 Operating Instructions ...
Страница 28: ...UltiMate 3000 Series AFC 3000 Automated Fraction Collector Page 22 Operating Instructions ...
Страница 48: ...UltiMate 3000 Series AFC 3000 Automated Fraction Collector Page 42 Operating Instructions ...
Страница 76: ...UltiMate 3000 Series AFC 3000 Automated Fraction Collector Page 70 Operating Instructions ...
Страница 78: ...UltiMate 3000 Series AFC 3000 Automated Fraction Collector Page 72 Operating Instructions ...
Страница 82: ...UltiMate 3000 Series AFC 3000 Automated Fraction Collector Page 76 Operating Instructions ...
Страница 86: ...UltiMate 3000 Series AFC 3000 Automated Fraction Collector Page 80 Operating Instructions Sample run ends here ...