Configuration
Thermo Foundation Instrument Configuration
Thermo Scientific
Accela PDA Detector Hardware Manual
29
3. Complete the configuration of the Accela PDA Detector as follows:
a. Double-click the
Accela PDA
button.
The Accela PDA Configuration dialog box appears (see
).
Figure 23.
Accela PDA Configuration dialog box (Thermo Foundation)
b. In the Stack Number box, type the number that the unit ID rotary switches are set to
on the back panel of the detector.
c.
In the Diode Array Scan Rate area, select the appropriate diode array scan rate for
your application.
The appropriate diode array scan rate depends on the baseline width (W
b
) of your
application’s chromatographic peaks:
d. Click
OK
to close the Accela PDA Configuration dialog box.
4. Complete the configuration of the other system devices, such as the pump and
autosampler.
5. Click
Done
to close the Instrument Configuration dialog box.
Baseline peak width (seconds)
Diode array scan rate (Hz)
W
b
0.5
80
0.5 < W
b
< 1
40
1
W
b
< 2
20
IMPORTANT
When you change the diode array scan rate, you must adjust the
light throughput to the diode array (see
“Adjusting the Light Throughput with
≥
≥