Instrument Driver Configuration
Foundation Instrument Configuration
74
Accela 600 and 1250 Pump Hardware Manual
Thermo Scientific
Adding Pump 2 to the Instrument Configuration
To add pump 2 to the instrument configuration for a dual-pump system
1. Ensure that pump 2 is connected to the data system computer.
2. Turn on pump 2.
3. From the computer desktop, choose
Start > All Programs > Thermo Foundation x.x >
Instrument Configuration
, where x.x is the installed version of Foundation.
The Unloading Device Drivers status box appears (see
), followed by
the Thermo Foundation Instrument Configuration application.
4. In the Configured Devices area, double-click the Accela
pump type
icon.
The Accela
pump type
dialog box appears.
5. Select the Pump 2 check box (see
Figure 64.
Accela 1250 Pump dialog box with the settings for a dual-pump system
6. Ensure that the Serial Number list displays the pump’s serial number and that the Status
readback displays Connected.
The serial number is listed on the pump’s back panel (see
).
7. If your Accela pump has a leak sensor, select the
Leak Sensor
Installed
check box.
8. Click
OK
to accept the configuration settings and close the Accela
pump type
dialog box.
9. Click
Done
to close the Instrument Configuration application.