2
Accela 600 Pump or Accela 1250 Pump Installation
Solvent Line Connections
Thermo Scientific
Accela 600 and 1250 Pump Hardware Manual
31
Connecting the High-Pressure Solvent Line to the Pump Outlet
Use the supplied precut, stainless steel tubing (0.005 in. ID × 1/16 in. OD) to connect the
pump to the autosampler:
• For the Accela Autosampler, use the 26 cm (10 in.) length of precut tubing in the Accela
System Kit.
• For the Accela Open Autosampler, use the 0.6 m (2 ft) length of precut tubing in the
Accela Open Autosampler Accessory Kit.
To connect the high-pressure solvent line to the LC pump outlet
1. Turn off the pump flow.
2. Using the supplied two-piece, stainless steel fitting, connect the precut, stainless steel
tubing to the pump outlet (dynamic mixer outlet [see
] or placeholder blind
module outlet).
Figure 17.
Dynamic mixer outlet
Tip
To identify the tubing ID, the precut tubing has a color-coded band:
• Red = 0.005 in.
• Black = 0.007 in.
• Blue = 0.010 in.
• Yellow = 0.020 in.
shows the proportioning valve design for pumps manufactured in
August 2011 or later.
Dynamic mixer outlet