Accela 600 Pump or Accela 1250 Pump Installation
Back Panel Connections
Thermo Scientific
Accela 600 and 1250 Pump Hardware Manual
23
Figure 11.
Adapter cable for an Accela detector (P/N 60157-63026)
To connect the contact closure cables to the Accela LC system modules
1. Connect the Accela Open Autosampler interconnect cable (P/N 60157-63024) to the
Accela Open Autosampler and the Accela 600 or 1250 Pump as follows:
a. Plug the DB15 connector into the Interface 1 receptacle on the back panel of the
Accela Open Autosampler.
b. Plug the 25-pin connector into the D-Sub25 receptacle on the back panel of the
Accela 600 or 1250 pump.
2. Connect the 2-pin connector labeled “TO 60157-63024” on the adapter cable to the
2-pin connector labeled “TO MSQ/LTQ/LXQ” on the Accela Open Autosampler
interconnect cable.
3. Connect the adapter cable to pins 1 through 8 on the back panel of the Accela detector.
shows the contact closure connections for an Accela LC system with an Accela
Open Autosampler, Accela 600 or 1250 Pump, and Accela detector.
Tip
For a dual-pump system, connect the 25-pin connector to the eluting pump.
Do not connect a contact closure cable to the loading pump.
For information about setting up the configuration for the device drivers, see
Chapter 4, “Instrument Driver Configuration.”
up the trigger signal for the loading pump (typically pump 2) and the pump
programs for both pumps, refer to the Accela user guide for your data system.