3
Hardware Upgrades
Dynamic Mixer Installation
42
Accela 600 and 1250 Pump Hardware Manual
Thermo Scientific
Connecting the Dynamic Mixer to the Pump’s Front Panel
To connect the dynamic mixer to the front panel of the pump
1. Using a 2.5 mm hex ball driver, remove the two screws (with washers) that secure the
blind outlet (placeholder) module to the front panel of the pump (see
). Set the
screws and washers aside. You use them in
to secure the dynamic mixer
to the pump.
Figure 27.
Back side of the pump’s front panel showing the screws (with washers) being
removed
Screws that secure the blind outlet (placeholder module)
to the front panel of the pump