Create protocol steps
This is the Create step view, which is where the actual protocol is programmed in a stepwise manner.
More details about the different functionalities are shown, see Table 2.
Note:
At a high-level, there are three parameters that can be set in a step, which are: magnet, rocker,
and pump. The setting of these should be considered for each step.
1
4
7
5
6
8
9
10
11
12
3
2
1
Start/end point of fluid lines
2
Engage/disengage magnet
3
Engage/disengage continous rocking
4
Start point of air
5
Add prompts
6
Rocker settings
7
Bubble sensor settings
8
Pump settings
9
Create next step by duplicating current
10
Navigate steps
11
Cancel protocol writing
12
Review all steps
Table 2 Features of the Create step view
Feature
Details
Magnet
Can be on or off.
When a duration is set for the magnet, it determines how long the rocker will be in the
specified position before moving to the next step.
Rocker
When off, a static angle can be set, when on, continuous rocking is specified.
Angle range: -30° to +30°
Speed range (cycles per minute): 1–30 CPM
The rocking can be symmetric (e.g. -10° to 10°) or “asymmetric” (e.g. -5° to 30°).
Chapter 3
Embedded graphical user interface (eGUI) user instructions
Manage protocols
3
54
CTS
™
DynaCellect
™
Magnetic Separation System User Guide
Содержание A55867