1. With the incubator ON, simultaneously press and hold the UP and
DOWN keys, then press the ENTER key. The letters will be shown in
the display. A number other than 96 might be shown in the display:
Choose the number appropriate for the Baud Rate desired.
3 = 300
6 = 600
12 = 1200
24 = 2400
48 = 4800
96 = 9600
Press the UP or DOWN key until the display shows your selection.
Press ENTER. The baud rate has been entered.
2. The display should change to the following. Again, the number in the
display might be different than 0.1, but it will have a decimal point.
Choose the number appropriate for the Status Update Time Interval
desired.
0.0 Minutes . . . . . . . . . . . . . . . . . . . . .No report
0.1 Minutes . . . . . . . . . . . . . . . . .Between reports
0.5 Minutes . . . . . . . . . . . . . . . . .Between reports
1.0 Minutes . . . . . . . . . . . . . . . . .Between reports
2.0 Minutes . . . . . . . . . . . . . . . .Between reports
5.0 Minutes . . . . . . . . . . . . . . . . .Between reports
10.0 Minutes . . . . . . . . . . . . . . . .Between reports
30.0 Minutes . . . . . . . . . . . . . . . .Between reports
60.0 Minutes . . . . . . . . . . . . . . . .Between reports
Press the UP or DOWN key until the setpoint display shows your
selection. Press ENTER. The value is entered.
Refrigerated Incubator
6-3
Thermo Scientific
Section 6
F
actor
y
Options
Set
U
p the Incubator for
Communications (cont.)