Aerosol-tight Applications
Checking the Aerosol-Tightness
5-4
75006476 LH-4000W
Thermo Scientific
•
Lubricate all seals lightly.
Always use the special grease 76003500 when lubricating the seals.
CAUTION
Prior to each use, the seals in the bucket are to be inspected in order to assure
that they are correctly seated and are not worn or damaged. Lubricate all seals lightly.
Damaged seals must be replaced immediately.
When loading the rotor, ensure that the rotor lid closes securely.
Damaged or clouded rotor caps must be replaced immediately.
Quick Test
CAUTION
This quick test is not suited for validating the aerosol tightness of
your bucket.
Check the seals and sealing surfaces of the bucket cap thoroughly.