7
Trouble Shooting
Troubleshooting by User
Thermo Scientific
Heraeus Multifuge X3F / X3FR
7-3
5. Reconnect the centrifuge once the power has been restored. Switch on the centrifuge. Press the
OPEN key
to have the door locks operative again.
Troubleshooting by User
WARNING
If you pull only one lever or you do not pressed the OPEN key
to have
the door locks operative again, the door might open while the rotor os still spinning.
If problems occur other than those listed in this table, the authorized customer service
representative must be contacted.
Failure message
Problem with centrifuge
Possible causes and cures
The centrifuge cannot be operated.
Overtemperature
The run does not start or the centrifuge runs
down without being braked.
Overheating in chamber.
Check the function of the refrigeration unit.
Clean the air inlet for the condenser.
Restart the centrifuge.
If an error message appears again, inform Customer Service.
The centrifuge cannot be operated.
Incorrect bucket ID
The run does not start or the centrifuge runs
down without being braked.
Bucket code is undefined for the rotor detected, check the set
points for the given bucket code.
Is it permitted to use the current bucket in the rotor currently
mounted?
Restart the centrifuge.
If an error message appears again, inform Customer Service.
The centrifuge cannot be operated.
Unapproved rotor
The run does not start or the centrifuge runs
down without being braked.
Rotor code is not in the rotor table.
Is it permitted to use the rotor currently mounted in this device?
Restart the centrifuge.
If an error message appears again, inform Customer Service.
The centrifuge cannot be operated.
Rotor ID failure
The run does not start or the centrifuge runs
down without being braked.
The rotor could not be identified.
Check to see if the rotor is properly installed.
Restart the centrifuge.
If an error message appears again, inform Customer Service.
Restart the centrifuge.
Centrifuge does not open
Lid blocked
The emergency lid release enables you to retrieve your samples.
If an error message appears again, inform Customer Service.
The centrifuge cannot be operated.
Motor overtemp.
The run does not start or the centrifuge runs
down without being braked.
Restart the centrifuge.
If an error message appears again, inform Customer Service.