3
AutoLock
TM
Rotor Installation
3-2
75003603 BIOShield
TM
1000A
Thermo Scientific
Rotor Installation
This rotor is equipped with an AutoLock
TM
-system.
This system is used to automatically lock the rotor to the centrifuge spindle. The rotor does not have
to be bolted onto the centrifuge spindle.
Proceed as follows:
1. Open the lid of the centrifuge and if necessary remove any dust, foreign objects or residue from
the chamber.
AutoLock
TM
and o-ring must be clean and undamaged.
Figure 3-1.
AutoLock
TM
2. Hold the rotor over the centrifuge spindle and let it slide slowly down the centrifuge spindle.
The rotor clicks automatically into place.
3. Check if the rotor is properly installed by lifting slightly on the handle. If the rotor can be pulled
up, then it must be reclamped to the centrifuge spindle.
CAUTION
Unapproved or incorrectly combined accessories can cause serious damage to
the centrifuge.
CAUTION
Do not force the rotor onto the centrifuge spindle.
If the rotor is very light, then it may be necessary to press it onto the centrifuge spindle
with a small amount of pressure.
WARNING
If the rotor cannot be properly locked in place after several attempts, then the
AutoLock
TM
is defective and you are not permitted to operate the rotor.
Check for any damage to the rotor, damaged rotors must not be used.
Keep the hub area clear of objects.
AutoLock
TM
O-ring
Содержание 75003603
Страница 3: ......
Страница 6: ...C Contents ii 75003603 BIOShieldTM 1000A Thermo Scientific...
Страница 22: ...3 AutoLockTM Removing the Rotor 3 4 75003603 BIOShieldTM 1000A Thermo Scientific...
Страница 30: ...5 Aerosol tight Applications Checking the Aerosol Tightness 5 4 75003603 BIOShieldTM 1000A Thermo Scientific...
Страница 36: ...6 Maintenance and Care Service of Thermo Fisher Scientific 6 6 75003603 BIOShieldTM 1000A Thermo Scientific...