background image

Thermo Scientific

 

BIOShield

TM

 1000A

Instruction

 Manual

50119946-

f

  •  0

7

 / 2020

Visit us online to register your warranty:

thermofisher.com/labwarranty

Содержание 75003603

Страница 1: ...Thermo Scientific BIOShieldTM 1000A Instruction Manual 50119946 f 07 2020 Visit us online to register your warranty thermofisher com labwarranty...

Страница 2: ...WEEE Conformity This product is subject to the regulations of the EU Waste Electrical Electronic Equipment WEEE Directive 2012 19 EU It is marked by the following symbol...

Страница 3: ......

Страница 4: ...rotor 75003603 Report No 59 08 G Report prepared for Thermo Fisher Issue Date 15th January 2009 Test Summary A Thermo Scientific 75003603 contained rotor Max speed 6 030 rpm was supplied by Thermo Fi...

Страница 5: ...ading 4 1 Before a Run 4 2 Proper Loading 4 2 Improper Loading 4 3 Maximum Loading 4 3 Cycle Counter 4 4 Chapter 5 Aerosol tight Applications 5 1 Basic Principles 5 2 Placing O ring 5 2 Fill Level 5 2...

Страница 6: ...C Contents ii 75003603 BIOShieldTM 1000A Thermo Scientific...

Страница 7: ...n the expiration of the seller s warranty Scope of Supply If any parts are missing please contact the nearest Thermo Fisher Scientific representative Precautions In order to ensure safe operation of t...

Страница 8: ...se pay particular attention to the following aspects Rotor installation Check that the rotor is locked properly into place before operating the centrifuge Always balance the samples Maximum sample den...

Страница 9: ...Thermo Scientific 75003603 BIOShieldTM 1000A 1 1 1 Rotor Specifications Contents Technical Data on page 1 2...

Страница 10: ...g at nmax C referred to ambient temperature of 23 25 C running time 120 minutes 12 12 8 Aerosol tight Tested by HPA Porton down UK yes yes yes Permissible temperature range autoclavable cycle number C...

Страница 11: ...C referred to ambient temperature of 23 25 C running time 120 minutes 12 12 8 Aerosol tight Tested by HPA Porton down UK yes yes yes Permissible temperature range autoclavable cycle number C 121 121 1...

Страница 12: ...nmax C referred to ambient temperature of 23 25 C running time 120 minutes 12 12 8 Aerosol tight yes yes yes Permissible temperature range autoclavable cycle number C 121 121 121 Tested by HPA Porton...

Страница 13: ...max C referred to ambient temperature of 23 25 C running time 120 minutes 12 12 8 Aerosol tight Tested by HPA Porton down UK yes yes yes Permissible temperature range autoclavable cycle number C 121 1...

Страница 14: ...nmax C referred to ambient temperature of 23 25 C running time 120 minutes 12 12 8 Aerosol tight Tested by HPA Porton down UK yes yes yes Permissible temperature range autoclavable cycle number C 121...

Страница 15: ...mple heating at nmax C referred to ambient temperature of 23 25 C running time 120 minutes 8 12 Aerosol tight yes yes Permissible temperature range autoclavable cycle number C 121 121 Tested by HPA Po...

Страница 16: ...70 60 Sample heating at nmax C referred to ambient temperature of 23 25 C running time 120 minutes 8 Aerosol tight Tested by HPA Porton down UK yes Permissible temperature range autoclavable cycle num...

Страница 17: ...Thermo Scientific 75003603 BIOShieldTM 1000A 2 1 2 Accessories Contents Rotor data on page 2 1 Accessories on page 2 1...

Страница 18: ...0050 50 50 PC Oak Ridge Tube 10 6 030 29x114 Included 10 PP Sealing 75003750 4 4 3114 0050 50 50 Teflon Oak Ridge Tube 2 6 030 29x114 Included 2 PP Sealing 75003750 4 4 45 Flat or Round Bottom Tube 3...

Страница 19: ...Thermo Scientific 75003603 BIOShieldTM 1000A 3 1 3 AutoLockTM Contents Rotor Installation on page 3 2 Removing the Rotor on page 3 3...

Страница 20: ...rifuge spindle The rotor clicks automatically into place 3 Check if the rotor is properly installed by lifting slightly on the handle If the rotor can be pulled up then it must be reclamped to the cen...

Страница 21: ...ndle with both hands and press against the green AutoLockTM button At the same time pull the rotor directly upwards with both hands and remove it from the centrifuge spindle Make sure not to jam the r...

Страница 22: ...3 AutoLockTM Removing the Rotor 3 4 75003603 BIOShieldTM 1000A Thermo Scientific...

Страница 23: ...mo Scientific 75003603 BIOShieldTM 1000A 4 1 4 Rotor Loading Contents Before a Run on page 4 2 Proper Loading on page 4 2 Improper Loading on page 4 3 Maximum Loading on page 4 3 Cycle Counter on page...

Страница 24: ...mpatibility chart on page B 1 5 Make sure the tubes or bottles do not touch the bucket lid 6 Check the rotor bolts and apply grease 75003786 before using for the first time 7 Check that each bucket or...

Страница 25: ...The safety system of the centrifuge requires that you do not overload the rotor There are two options available for centrifuging samples whose weight including adapter exceeds the maximum permissible...

Страница 26: ...les is given in the rotor table in the rotor specification section The maximum number of cycles for buckets is marked on the buckets themselves Service life examples 750 4987 770 4922 Actual Load g bu...

Страница 27: ...Thermo Scientific 75003603 BIOShieldTM 1000A 5 1 5 Aerosol tight Applications Contents Basic Principles on page 5 2 Fill Level on page 5 2 Checking the Aerosol Tightness on page 5 2...

Страница 28: ...tube only 2 3 of the rated level Checking the Aerosol Tightness The aerosol tightness testing of the rotors and buckets depend on the microbiological test process in accordance with the EN 61010 2 020...

Страница 29: ...ater resulting in excess pressure Do not apply pressure to the lid when doing so Leaks can be detected by escaping water or the sound of escaping gas Replace the seals if you detect any leaks Then rep...

Страница 30: ...5 Aerosol tight Applications Checking the Aerosol Tightness 5 4 75003603 BIOShieldTM 1000A Thermo Scientific...

Страница 31: ...OShieldTM 1000A 6 1 6 Maintenance and Care Contents Cleaning intervalls on page 6 2 Cleaning on page 6 2 Disinfection on page 6 3 Decontamination on page 6 4 Autoclaving on page 6 5 Service of Thermo...

Страница 32: ...f the parts Use only disinfectants with a pH of 6 8 Dry aluminum parts off with a soft cloth After cleaning treat the entire surface of aluminum parts with corrosion protection oil 70009824 Also treat...

Страница 33: ...ectious material has spilled during centrifugation Use a sprayer whenever possible so that all surfaces are covered evenly The rotor chamber and the rotor should be treated preferably with a neutral d...

Страница 34: ...h a cloth or in a warm air cabinet at a maximum temperature of 50 C 11 After cleaning treat the entire surface of aluminum parts with corrosion protection oil 70009824 Also treat the cavities with oil...

Страница 35: ...the bolt of the swing out rotor with bolt grease 75003786 Autoclaving 1 Before autoclaving clean rotor and accessories and described above 2 Place the rotor on a flat surface Rotors and adapter can b...

Страница 36: ...6 Maintenance and Care Service of Thermo Fisher Scientific 6 6 75003603 BIOShieldTM 1000A Thermo Scientific...

Страница 37: ...199 1100 8 2 17 8 111 241 1200 8 2 17 8 132 287 1300 8 2 17 8 155 336 1400 8 2 17 8 180 390 1500 8 2 17 8 206 448 1600 8 2 17 8 235 509 1700 8 2 17 8 265 575 1800 8 2 17 8 297 645 1900 8 2 17 8 331 71...

Страница 38: ...17 8 1541 3345 4200 8 2 17 8 1617 3510 4300 8 2 17 8 1695 3680 4400 8 2 17 8 1775 3853 4500 8 2 17 8 1856 4030 4600 8 2 17 8 1940 4211 4700 8 2 17 8 2025 4396 4800 8 2 17 8 2112 4585 4900 8 2 17 8 220...

Страница 39: ...S S S U S S S S M S S S S S S S S S M U U S S Formic Acid 100 S M U U U S M U U S S U S U S U Ammonium Acetate S S U S S S S S S S S S S U S S S S S S S S S S Ammonium Carbonate M S U S S S S S S S S...

Страница 40: ...S S U S U S U S S S M S S S S S S Dimethylsulfoxide S S U U S S S S U S S U S U U S S U U S S S S U U Dioxane M S U U S S M M S U U S U M U U M M M U S S S S U U Ferric Chloride U U S M S M S S S S M...

Страница 41: ...Hydroxide conc U U M U M M S S U M U U U S M M U U U U Potassium Permanganate S S S S S S S S S U S S S M S M S U S S M S U S Calcium Chloride M U S S S S S S S S S S S S M S S S S S S S M S S S Calc...

Страница 42: ...S S S S S S M S S S S S S M S S S S S S S S S S Nickel Salts U S S S S S S S S S S S S S S S S S S M S S S Oils Petroleum S S S S U S S S S U U M S M U U S S S U S S S S Oils Other S S S M S S S S U S...

Страница 43: ...U U U S U S S U U Toluene S S U U S S M U S U U S U U U S U M U U U S U S U U M Trichloroacetic Acid U U U S S U M S U S U U S M M S S U U S U U U M U Trichloroethane S U M U U S U U U U U U U U U S U...

Страница 44: ...ials under the stress of centrifugation when in doubt we recommend pretesting sample lots Polyethyleneterephthalate Key S Satisfactory M Moderate attack may be satisfactory for use in centrifuge depen...

Страница 45: ...g 6 2 Cleaning Agents 6 3 Contact Information C 1 Cycle Counter 4 4 D Decontamination 6 4 Disinfection 6 3 F Fill Level 5 2 Technical Data 1 2 I Improper Loading 4 3 M Maintenance 6 1 Maximum Loading...

Страница 46: ...consult your local sales represen tative for details Shown pictures within the manual are examples and may differ considering the set parameters and language Pictures of the user interface within the...