Operation
Integration Time
3-12
Model 5028
i
Instruction Manual
Thermo Fisher Scientific
The Custom Ranges screen is used to define the custom ranges.
The display shows the current custom range. The next line of the display
is used to set the range. For more information about selecting ranges,
” above.
l
In the Main Menu, choose Range > Set Custom Ranges >
Custom
Range 1, 2,
or
3
.
CUSTOM RANGE 1:
CURRENTLY: 555.6
SET TO: 000555.7 ?
MOVE CURSOR
CHANGE VALUE SAVE
RANGE AVG DIAGS ALARM
Integration time menu allows the operator to select the measurement of
time interval for the channel A or channel B. To adjust the integration
time interval, Integration Time ensure the measured values are
calculated by the average value, and the average value is measured as the
output of the measured value.Integration time interval can be selected
from the 6 values that are preset (15, 20, 30, 40, 45 and 60 minutes).
Note
Integration Time B settings please refer to Integration Time A as
following description.
s
l
In the Main Menu, choose Integration Time>
Integration Time A
or Integration Time B
.
INTEGRATION TIME:
>INTEGRATION TIME A
INTEGRATION TIME B
RANGE AVG DIAGS ALARM
INTEGRATION TIME A:
CURRENTLY: 20 MIN
SET TO: 15 MIN ?
CHANGE VALUE
SAVE VALUE
RANGE AVG DIAGS ALARM
The 24-Hour Average screen defines a start time during which
concentration measurement is averaged over a daily 24-hours period.
The average concentration of the primary readings are calculated for that
time period and averaged out. A new 24-hour average concentration is
displayed once per day. The display shows the current 24-hour average
PM/PM
b
concentration. The next line of the display is used to change
the 24-hour averaging time.
Custom Ranges
Integration Time
24-Hour Average
Содержание 5028i
Страница 1: ...Model 5028i Instruction Manual Continuous Particulate Monitor CIC0001582 April 2018...
Страница 2: ......
Страница 3: ...Model 5028i Instruction Manual Continuous Particulate Monitor Part Number CIC0001582 8Apr2018...
Страница 6: ...WEEE Compliance Thermo Fisher Scientific...
Страница 12: ...About This Manual WEEE Symbol vi Model 5028i Instruction Manual Thermo Fisher Scientific...
Страница 36: ......
Страница 142: ......
Страница 154: ......
Страница 220: ......
Страница 222: ...System Description Hardware 8 2 Model 5028i Instruction Manual Thermo Fisher Scientific Figure 8 1 Hardware Components...
Страница 232: ......
Страница 328: ......
Страница 329: ......