Operation
Range Menu
3-10
Model 5028
i
Instruction Manual
Thermo Fisher Scientific
CONC UNITS:
CURRENTLY: mg/m3
SET TO:
3
µ
g/m3 ?
CHANGE VALUE SAVE
RANGE AVG DIAGS ALARM
The Range screen defines the concentration range of the analog outputs.
For example, PM/PM
b
range of 0–100 µg/m
3
sets the analog output to
concentrations between 0 and 100 µg/m
3
by default.
Note
PM
b
Range settings please refer to PM Range as following
description.
s
The display shows the current PM/PM
b
range. The next line of the
display is used to change the range.
lists the standard ranges.
l
In the Main Menu, choose Range >
PM Range or PM
b
Range
.
PM RANGE:
CURRENTLY: 100.0
SET TO: 1000 ?
CHANGE VALUE
SAVE VALUE
RANGE AVG DIAGS ALARM
Table 3–2.
Standard Ranges
mg/m
3
µ
g/m
3
0.1
100
1.0
1000
2.0
2000
3.0
3000
5.0
5000
10.00
10000
C1
C1
C2
C2
C3
C3
C1, C2, and C3 are custom ranges. For more information about custom
ranges, see “Set Custom Ranges” below.
The analog outputs are arranged on the rear panel connector as shown
in
for channels and pin connections.
Range
Содержание 5028i
Страница 1: ...Model 5028i Instruction Manual Continuous Particulate Monitor CIC0001582 April 2018...
Страница 2: ......
Страница 3: ...Model 5028i Instruction Manual Continuous Particulate Monitor Part Number CIC0001582 8Apr2018...
Страница 6: ...WEEE Compliance Thermo Fisher Scientific...
Страница 12: ...About This Manual WEEE Symbol vi Model 5028i Instruction Manual Thermo Fisher Scientific...
Страница 36: ......
Страница 142: ......
Страница 154: ......
Страница 220: ......
Страница 222: ...System Description Hardware 8 2 Model 5028i Instruction Manual Thermo Fisher Scientific Figure 8 1 Hardware Components...
Страница 232: ......
Страница 328: ......
Страница 329: ......