Installation
Acceptance Testing and Startup Procedures
Thermo Fisher Scientific
Model 5028
i
Instruction Manual
2-3
After acceptance testing, a completed monitoring installation will
require final volumetric flow rate verification.
Prior to installing the Model 5028
i
, you should perform the acceptance
testing procedures. These tests are conducted to evaluate the out-of-
box performance of the instrument, perform any necessary calibrations
prior to final site installation, and familiarize the user with the menu
structure.
Use the following procedure to perform the bench acceptance test.
Equipment Required:
Model 5028
i
Power Cord
Ambient Temperature/RH Cable Assembly
Vacuum Pump Assembly
Flow Adapter Assembly
NIST-traceable Thermometer
NIST-traceable Hygrometer
NIST-traceable Barometer
NIST-traceable Manometer(s)
NIST-traceable Volumetric Flow Transfer Standard
(
Do not use the heated sample tube assembly at this point
.)
1. After unpacking the instrument place the Model 5028
i
and
accessories onto a table or bench located within a stable indoor
environment. Open the front door and inspect the filter tapes. Leave
the front door unlocked until after acceptance testing has been
completed.
Note
Prior to continuing with acceptance testing, the instrumentation
and accessories should have sufficient time to equilibrate to room
temperature due to temperature variations during shipping and/or
storage.
s
2. Following a sufficient equilibration period, connect the power cord
to the rear of the instrument and to a properly grounded power
supply.
Note
If the Model 5028
i
turns ON at this point, set the power switch
to the OFF position and then proceed.
s
3. Connect the green vacuum tube to the pump intake and to one end
of the blue fitter. Connect a short green vacuum tube to the other
end of blue fitter and to Tee-union. By using transparent tube, port
Acceptance Test
Содержание 5028i
Страница 1: ...Model 5028i Instruction Manual Continuous Particulate Monitor CIC0001582 April 2018...
Страница 2: ......
Страница 3: ...Model 5028i Instruction Manual Continuous Particulate Monitor Part Number CIC0001582 8Apr2018...
Страница 6: ...WEEE Compliance Thermo Fisher Scientific...
Страница 12: ...About This Manual WEEE Symbol vi Model 5028i Instruction Manual Thermo Fisher Scientific...
Страница 36: ......
Страница 142: ......
Страница 154: ......
Страница 220: ......
Страница 222: ...System Description Hardware 8 2 Model 5028i Instruction Manual Thermo Fisher Scientific Figure 8 1 Hardware Components...
Страница 232: ......
Страница 328: ......
Страница 329: ......