MODBUS Protocol
MODBUS Parameters Supported
C-8
Model 5028
i
Instruction Manual
Thermo Fisher Scientific
Request
Field Name
(Hex)
Function
05
Output Address Hi
00
Output Address Lo
05
Output Value Hi
FF
Output Value Lo
00
Response
Field Name
(Hex)
Function
05
Output Address Hi
00
Output Address Lo
05
Output Value Hi
FF
Output Value Lo
00
lists the MODBUS addresses supported
for the Model 5028
i
.
IMPORTANT NOTE
The addresses in the following tables are Protocol
Data Unit (PDU) addresses. Verify the coil number on your MODBUS
master to ensure that it matches the coil number on the instrument.
s
Note
Coil status 1 indicates active state.
s
Table C–1.
Read Coils for 5028
i
Coil Number
Status
0
INVALID
1
LOCAL/REMOTE
2
SERVICE
3
[NOT USED]
4
[NOT USED]
5
CONC ALARM
6
INST ALARMS
7
DET ALARMS
8
NEPH ALARMS
MODBUS Parameters
Supported
Содержание 5028i
Страница 1: ...Model 5028i Instruction Manual Continuous Particulate Monitor CIC0001582 April 2018...
Страница 2: ......
Страница 3: ...Model 5028i Instruction Manual Continuous Particulate Monitor Part Number CIC0001582 8Apr2018...
Страница 6: ...WEEE Compliance Thermo Fisher Scientific...
Страница 12: ...About This Manual WEEE Symbol vi Model 5028i Instruction Manual Thermo Fisher Scientific...
Страница 36: ......
Страница 142: ......
Страница 154: ......
Страница 220: ......
Страница 222: ...System Description Hardware 8 2 Model 5028i Instruction Manual Thermo Fisher Scientific Figure 8 1 Hardware Components...
Страница 232: ......
Страница 328: ......
Страница 329: ......