C-Link Protocol Commands
Measurement Configuration
B-44
Model 5028
i
Instruction Manual
Thermo Fisher Scientific
set range a/b
selection
This command selects the PM ranges, according to
. The
following example sets the PM range to 10 mg/m
3
.
Send:
set range a 5
Receive:
set range a 5 ok
Send:
set range b 5
Receive:
set range b 5 ok
Table B–5.
Standard Ranges
Selection
m
g/m3
mg/m3
0
100
0.1
1
1000
1.0
2
2000
2.0
3
3000
3.0
4
5000
5.0
5
10000
10.0
6
C1
C1
7
C2
C2
8
C3
C3
rh th a/b
This command reports the current heater RH threshold for the sample.
The following example reports that the RH threshold is 30%.
Send:
rh th a
Receive:
rh th a 30
%
Send:
rh th b
Receive:
rh th b 30
%
set rh th a/b
value
This command sets the heater RH threshold for the sample to
value
,
where
value
is a floating-point number representation of RH threshold
between 30 to 95%. The following example sets the RH threshold to
35%.
Send:
set rh th a 35
Receive:
set rh th a 35 ok
Send:
set rh th b 35
Receive:
set rh th b 35 ok
Содержание 5028i
Страница 1: ...Model 5028i Instruction Manual Continuous Particulate Monitor CIC0001582 April 2018...
Страница 2: ......
Страница 3: ...Model 5028i Instruction Manual Continuous Particulate Monitor Part Number CIC0001582 8Apr2018...
Страница 6: ...WEEE Compliance Thermo Fisher Scientific...
Страница 12: ...About This Manual WEEE Symbol vi Model 5028i Instruction Manual Thermo Fisher Scientific...
Страница 36: ......
Страница 142: ......
Страница 154: ......
Страница 220: ......
Страница 222: ...System Description Hardware 8 2 Model 5028i Instruction Manual Thermo Fisher Scientific Figure 8 1 Hardware Components...
Страница 232: ......
Страница 328: ......
Страница 329: ......