Servicing
Proportional Valve Replacement
Thermo Fisher Scientific
Model 5028
i
Instruction Manual
7-27
Unhook Cable
Unfasten screws (2)
Figure 7–14.
Replacing the Photo Interrupt Board
Use the following procedure to replace the proportional valve (
).
Equipment Required:
Proportional valve assembly
Hex driver, 3/32-inch
Equipment Damage
Some internal components can be damaged by
small amounts of static electricity. A properly grounded antistatic wrist
strap must be worn while handling any internal component.
s
1. Turn the instrument OFF, unplug the power cord, and remove the
cover.
2. Refer to “
Removing the Measurement Case Assembly and Rear
” in this chapter to lower the partition panel, then proceed to
the next step below.
3. Disconnect connector from the measurement interface board.
4. Disconnect plumbing.
5. Remove four mounting screws and remove proportional valve
assembly.
6. Replace the proportional valve assembly by following the previous
steps in reverse.
7. Perform a leak test as described in the “
chapter.
Proportional Valve
Replacement
Содержание 5028i
Страница 1: ...Model 5028i Instruction Manual Continuous Particulate Monitor CIC0001582 April 2018...
Страница 2: ......
Страница 3: ...Model 5028i Instruction Manual Continuous Particulate Monitor Part Number CIC0001582 8Apr2018...
Страница 6: ...WEEE Compliance Thermo Fisher Scientific...
Страница 12: ...About This Manual WEEE Symbol vi Model 5028i Instruction Manual Thermo Fisher Scientific...
Страница 36: ......
Страница 142: ......
Страница 154: ......
Страница 220: ......
Страница 222: ...System Description Hardware 8 2 Model 5028i Instruction Manual Thermo Fisher Scientific Figure 8 1 Hardware Components...
Страница 232: ......
Страница 328: ......
Страница 329: ......