Servicing
Detector Assembly Replacement
7-28
Model 5028
i
Instruction Manual
Thermo Fisher Scientific
Figure 7–15.
Replacing the Proportional Valve
Use the following procedure to replace the detector assembly (
).
Equipment Required:
Detector assembly
Hex driver, 4 mm
Safety glasses
WARNING
The detector assembly should be replaced by a qualified
technician knowledgeable in dealing with radiation precautions.
By removing the detector assembly, the C-14 radioactive source is
partially exposed. Safety glassed must be worn during this replacement
procedure. The amount of C-14 meets the USNRC regulations as an
exempt amount of radioactive source <100 µCi.
s
The service procedures in this manual are restricted to qualified
representatives.
s
If the equipment is operated in a manner not specified by the
manufacturer, the protection provided by the equipment may be
impaired.
s
CAUTION
The detector window is very fragile. Handle with great care
and do not wipe or touch the window. Furthermore, do not touch the
solder cable connection (electrode) with your bare fingers. The oils from
your skin can damage the detector.
s
Detector Assembly
Replacement
Содержание 5028i
Страница 1: ...Model 5028i Instruction Manual Continuous Particulate Monitor CIC0001582 April 2018...
Страница 2: ......
Страница 3: ...Model 5028i Instruction Manual Continuous Particulate Monitor Part Number CIC0001582 8Apr2018...
Страница 6: ...WEEE Compliance Thermo Fisher Scientific...
Страница 12: ...About This Manual WEEE Symbol vi Model 5028i Instruction Manual Thermo Fisher Scientific...
Страница 36: ......
Страница 142: ......
Страница 154: ......
Страница 220: ......
Страница 222: ...System Description Hardware 8 2 Model 5028i Instruction Manual Thermo Fisher Scientific Figure 8 1 Hardware Components...
Страница 232: ......
Страница 328: ......
Страница 329: ......