Servicing
I/O Expansion Board (Optional) Replacement
Thermo Fisher Scientific
Model 5028
i
Instruction Manual
7-23
Use the following procedure to replace the optional I/O expansion
board (
Equipment Required:
I/O expansion board
Nut driver, 3/16-inch
Equipment Damage
Some internal components can be damaged by
small amounts of static electricity. A properly grounded antistatic wrist
strap must be worn while handling any internal component.
s
1. Turn the instrument OFF, unplug the power cord, and remove the
cover.
2. Unplug the I/O expansion board cable from the EXPANSION BD
connector on the motherboard.
3. Remove the two standoffs holding the I/O expansion board
connector to the rear panel (
).
4. Pop the board off of the mounting studs and remove the board.
5. To install the I/O expansion board, follow the previous steps in
reverse.
6. Calibrate the analog current outputs and analog voltage inputs as
defined earlier in this chapter.
Figure 7–11.
Replacing the I/O Expansion Board (Optional)
I/O Expansion
Board (Optional)
Replacement
Содержание 5028i
Страница 1: ...Model 5028i Instruction Manual Continuous Particulate Monitor CIC0001582 April 2018...
Страница 2: ......
Страница 3: ...Model 5028i Instruction Manual Continuous Particulate Monitor Part Number CIC0001582 8Apr2018...
Страница 6: ...WEEE Compliance Thermo Fisher Scientific...
Страница 12: ...About This Manual WEEE Symbol vi Model 5028i Instruction Manual Thermo Fisher Scientific...
Страница 36: ......
Страница 142: ......
Страница 154: ......
Страница 220: ......
Страница 222: ...System Description Hardware 8 2 Model 5028i Instruction Manual Thermo Fisher Scientific Figure 8 1 Hardware Components...
Страница 232: ......
Страница 328: ......
Страница 329: ......