Servicing
Detector Amplifier Board Calibration
7-12
Model 5028
i
Instruction Manual
Thermo Fisher Scientific
3. Remove the one screw holding the detector amplifier mounting
bracket and slide out the detector amplifier assembly.
Figure 7–5.
Replacing the Detector Amplifier Assembly
4. Install the new detector amplifier assembly by following the previous
steps in reverse.
5. Be sure to the new detector amplifier assembly connect to the right
measurement head. The detector amplifier A couldn’t be replaced by
detector amplifier B, and vice versa.
6. Calibrate the detector amplifier board. Refer to the “
” procedure that follows.
Note
If Service Mode is not displayed, refer to “
, then return to the beginning of this step.
s
7. Calibrate the instrument. Refer to the “
manual.
Calibrate the detector amplifier after replacing the detector amplifier
board.
Note
This adjustment should only be performed by an instrument
service technician.
s
1. Let the instrument sample zero air for about 90 minutes.
Detector Amplifier
Board Calibration
Содержание 5028i
Страница 1: ...Model 5028i Instruction Manual Continuous Particulate Monitor CIC0001582 April 2018...
Страница 2: ......
Страница 3: ...Model 5028i Instruction Manual Continuous Particulate Monitor Part Number CIC0001582 8Apr2018...
Страница 6: ...WEEE Compliance Thermo Fisher Scientific...
Страница 12: ...About This Manual WEEE Symbol vi Model 5028i Instruction Manual Thermo Fisher Scientific...
Страница 36: ......
Страница 142: ......
Страница 154: ......
Страница 220: ......
Страница 222: ...System Description Hardware 8 2 Model 5028i Instruction Manual Thermo Fisher Scientific Figure 8 1 Hardware Components...
Страница 232: ......
Страница 328: ......
Страница 329: ......