C-Link Protocol Commands
I/O Configuration
B-40
Model 49
i
Instruction Manual
Thermo Fisher Scientific
Send:
set analog iout range 4 1
Receive:
set analog iout range 4 1 ok
analog vin
channel
This command retrieves the analog voltage input
channel
data, both the
calculated value and the actual voltage. In the example that follows, the
“calculated” value of channel 1 is 75.325 degrees F, volts are 2.796. This
command responds with “feature not enabled” if the I/O expansion board is
not detected.
Send:
analog vin 1
Receive:
analog vin 1 75.325 2.796
analog vout range
channel
This command reports the analog voltage output
channel
range, where
channel
is between 1 and 6 inclusive, according to
.
Send:
analog vout range 2
Receive:
analog vout range 2 3
set analog vout range
channel range
This command sets analog voltage output
channel
to the range, where
channel
is between 1 and 6 inclusive, and
range
is set according to
. The example that follows sets channel 2 to the 0-10 V range.
Send:
set analog vout range 2 3
Receive:
set analog vout range 2 3 ok
Table B-9.
Analog Current Output Range Values
Range
Output Range
1
0-20 mA
2
4-20 mA
0 [cannot be set to this, but may report]
Undefined
Table B-10.
Analog Voltage Output Range Values
Range
Output Range
1
0-1 V
2
0-100 mV
3
0-10 V
4
0-5 V
0 [cannot be set to this, but may report]
Undefined
Содержание 49i
Страница 1: ...Model 49i Instruction Manual UV Photometric O3 Analyzer Part number 102434 00 10Jan2008 ...
Страница 4: ......
Страница 6: ......
Страница 14: ...Contents xiv Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 16: ...Figures xvi Model 49i Instruction Manual Thermo Fisher Scientific Flag Status B 11 ...
Страница 36: ...Installation Startup 2 10 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 162: ...Preventive Maintenance Ozone Scrubber Test 5 12 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 184: ...Troubleshooting Service Locations 6 22 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 244: ...Optional Equipment Mounting Options 9 6 Model 49i Instruction Manual Thermo Fisher Scientific Figure 9 2 Bench Mounting ...