C-Link Protocol Commands
Communications Configuration
Thermo Fisher Scientific
Model 49
i
Instruction Manual
B-37
set baud
rate
rate
= | 1200 | 2400 | 4800 | 9600 | 19200 | 38400 | 57600 | 115200 |
This command sets the instrument baud
rate
. The example that follows sets
the instrument’s baud rate to 115200.
Note
After the command is sent, the baud rate of the sending device must
be changed to agree with the instrument.
Send:
set baud 115200
Receive:
set baud 115200 ok
dhcp
This command reports the current state of use of DHCP on or off. DHCP
is used to assign an IP address to the analyzer automatically. The example
that follows shows that DHCP is on.
Send:
dhcp
Receive:
dhcp on
set dhcp
onoff
These commands enables and disables the DHCP service by either
on
or
off
.
Changes to this parameter will only take effect when the analyzer is powered
up. The example that follows sets the DHCP service on.
Note
When DHCP is set to on, the user-supplied addr gw, addr dns, addr
ip, and addr nm parameters are not used.
Send:
set dhcp on
Receive:
set dhcp on ok
format
This command reports the current reply termination format. The example
that follows shows that the reply format is 00, which means reply with no
checksum, according to
Send:
format
Receive:
format 00
set format
format
This command sets the reply termination
format
, where
format
is set
according to
. The example that follows sets the reply termination
format to checksum.
Send:
set format 01
Receive:
set format 01 ok
Содержание 49i
Страница 1: ...Model 49i Instruction Manual UV Photometric O3 Analyzer Part number 102434 00 10Jan2008 ...
Страница 4: ......
Страница 6: ......
Страница 14: ...Contents xiv Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 16: ...Figures xvi Model 49i Instruction Manual Thermo Fisher Scientific Flag Status B 11 ...
Страница 36: ...Installation Startup 2 10 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 162: ...Preventive Maintenance Ozone Scrubber Test 5 12 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 184: ...Troubleshooting Service Locations 6 22 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 244: ...Optional Equipment Mounting Options 9 6 Model 49i Instruction Manual Thermo Fisher Scientific Figure 9 2 Bench Mounting ...