C-Link Protocol Commands
Communications Configuration
B-38
Model 49
i
Instruction Manual
Thermo Fisher Scientific
where xxxx = 4 hexadecimal digits that represent the sum of all the
characters (bytes) in the message
host name
This command reports the host name string.
Send:
host name
Receive:
host name ISERIES
set host name
string
This command sets the host name
string
, where
string
is 1-13 alphanumeric
characters.
Send:
set host name analyzer01
Receive:
set host name analyzer01 ok
instr name
This command reports the instrument name.
Send:
instr name
Receive:
instr name
o3 Analyzer
03 Analyzer
instrument id
This command reports the instrument id.
Send:
instrument id
Receive:
instrument id 49
set instrument id
value
This command sets the instrument id to
value,
where
value
is a decimal
number between 0 and 127 inclusive.
Note
Sending this command via RS-232 or RS-485 will require the host to
use the new id for subsequent commands.
Send:
set instrument id 50
Receive:
set instrument id 50 ok
Table B-8.
Reply Termination Formats
Format
Reply Termination
00
<CR>
01
<NL> sum xxxx <CR>
Содержание 49i
Страница 1: ...Model 49i Instruction Manual UV Photometric O3 Analyzer Part number 102434 00 10Jan2008 ...
Страница 4: ......
Страница 6: ......
Страница 14: ...Contents xiv Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 16: ...Figures xvi Model 49i Instruction Manual Thermo Fisher Scientific Flag Status B 11 ...
Страница 36: ...Installation Startup 2 10 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 162: ...Preventive Maintenance Ozone Scrubber Test 5 12 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 184: ...Troubleshooting Service Locations 6 22 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 244: ...Optional Equipment Mounting Options 9 6 Model 49i Instruction Manual Thermo Fisher Scientific Figure 9 2 Bench Mounting ...