C-Link Protocol Commands
Measurement Configuration
B-30
Model 49
i
Instruction Manual
Thermo Fisher Scientific
custom
range
range
= | 1 | 2 | 3 |
This command reports the user-defined value of custom
range
1, 2, or 3.
The example that follows reports that custom range 1 is defined to 55.0
ppm.
Send:
custom 1
Receive:
custom 1 5500E-2 ppm
set custom
range value
set custom 1
value
set custom
2
value
set custom 3
value
set custom 1 range
value
set custom
2 range
value
set custom 3 range
value
These commands are used to set the maximum concentration for any of the
three custom
ranges
1, 2, or 3 to range
value
, where
value
is a floating-point
number representing concentration in ppm or mg/m
3
. The example that
follows sets the custom 1 range to 55.5 ppm.
Send:
set custom 1 range 55.5
Receive:
set custom 1 range 55.5 ok
range mode
This command reports the current range mode.
Send:
range mode
Receive:
range mode single
set range mode
mode
This command sets the current range
mode
to single, dual, or auto. The
example that follows sets the range mode to single.
Send:
set range mode single
Receive:
set range mode single ok
gas mode
This command reports the current mode of sample, zero, or span. The
example that follows reports that the gas mode is sample.
Send:
gas mode
Receive:
gas mode sample
Содержание 49i
Страница 1: ...Model 49i Instruction Manual UV Photometric O3 Analyzer Part number 102434 00 10Jan2008 ...
Страница 4: ......
Страница 6: ......
Страница 14: ...Contents xiv Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 16: ...Figures xvi Model 49i Instruction Manual Thermo Fisher Scientific Flag Status B 11 ...
Страница 36: ...Installation Startup 2 10 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 162: ...Preventive Maintenance Ozone Scrubber Test 5 12 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 184: ...Troubleshooting Service Locations 6 22 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 244: ...Optional Equipment Mounting Options 9 6 Model 49i Instruction Manual Thermo Fisher Scientific Figure 9 2 Bench Mounting ...