Optional Equipment
PTFE Particulate Filter
Thermo Scientific
43iQ Instruction Manual
9-23
A 5-10 micron pore size, two-inch diameter PTFE element is available for
the 43iQ. This filter should be installed just prior to the
SAMPLE
bulkhead.
When using a filter, all calibrations and span checks must be performed
through the filter.
The 43iQ ships standard with a capillary to maintain the instrument flow
rate at approximately 0.5 LPM. An optional capillary is available that will
maintain the instrument flow at approximately 1.0 LPM. This increased
flow may be desirable when a reduced instrument response time is desired.
To run the higher flow rate, replace the instrument capillary with a 0.02
size capillary. (Refer to “43iQ Measurement Case” in the
iQ Series Gas
Analyzer Spare Parts Catalog
for part number.)
From the Home Screen, press Configuration>
High Flow
to enable the
high flow option.
PTFE Particulate
Filter
High Flow Option
Содержание 43iQ
Страница 1: ...43iQ Instruction Manual Sulfur Dioxide Analyzer 117568 00 1Sep2021 ...
Страница 2: ......
Страница 8: ......
Страница 18: ......
Страница 34: ...Operation Instrument Display 3 8 43iQ Instruction Manual Thermo Scientific ...
Страница 162: ......
Страница 166: ...Maintenance Pump Rebuilding 5 4 43iQ Instruction Manual Thermo Scientific Figure 5 2 Pump Rebuilding ...
Страница 170: ......
Страница 180: ......
Страница 188: ...Servicing Fan Replacement 7 8 43iQ Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan ...
Страница 228: ......
Страница 238: ......
Страница 262: ......
Страница 270: ......
Страница 282: ......
Страница 283: ......