Servicing
Pump Replacement
Thermo Scientific
43iQ Instruction Manual
7-19
Use the following procedure to replace the pump.
Equipment required:
Phillips drive, #1 and #2
1.
Turn instrument OFF, unplug power cord, and remove the cover
2.
Unplug pump cable from step pol board J7.
3.
Twist opposite direction to unlock tube clamps.
Note
Push in tube clamp to lock.
▲
4.
Disconnect tubing from pump.
5.
Unfasten two captive hardware.
6.
Slide pump left until keyway meets opening.