Section 6
Maintenance
Warning
Disconnect plug from electrical outlet before attempting any
maintenance or repair of the unit.
s
Note
The shaking mechanism is equipped with sealed ball bearings which
do not require further lubrication or adjustment.
s
Note
Make no attempt to service or repair a Thermo Scientific product
under warranty before consulting your distributor. After the warranty
period, such consultation is still advised, especially when the repair may be
technically sophisticated or difficult.
If assistance is needed beyond what the distributor can provide, please call
Technical Services department. No merchandise, however, should be
returned without prior approval.
Wash the exterior of the unit with a soft cloth using a solution of mild
soap and water, rinse off with clean water and dry thoroughly.
Suggested with every 3 months of constant use
Any internal adjustments or repairs must be performed by a qualified
service representative.
Remove the platform by loosening 4, thumbscrews in the platform center.
Remove the sheet metal panel (9 screws) under the platform to expose the
belt and interior parts. Inspect the drive belt for wear. Order a replacement
if necessary.
Incubated and Refrigerated Shakers 6-1
Thermo Scientific