Calibration
Equipment Required
Thermo Scientific
42iQD Instruction Manual
4-3
Fixed bed reactors are commonly used in the last step of zero air generation
to remove the remaining contaminants by either further reaction or
lists materials that can be effective in removing
contaminants.
Table 4–1.
Scrubbing Materials
To Remove
Use
NO
Soda-Lime (6-12 mesh), Purafil
Hydrocarbons
Molecular Sieve (4A), Activated Charcoal
O
3
and NO
2
Activated Charcoal
A gas phase titrator (GPT), such as is included in the Thermo Scientific
146iQ Multi-gas Calibrator
,
is used to generate NO
2
concentrations from
shows the suggested placement of the
component parts of a gas phase titration apparatus.
Equipment Damage
All connections between components in the system
should be made with glass, PTFE, or other non-reactive material.
▲
The airflow controllers should be devices capable of maintaining constant
airflows within
±
2% of the required flow rate. The NO flow controller
should be capable of maintaining constant NO flows within
±
2% of the
required flow rate.
The pressure regulator for the standard NO cylinder must have a non-
reactive diaphragm and internal parts, and a suitable delivery pressure.
Scrubbing
Gas Phase Titrator
Flow Controllers
Pressure Regulator
Содержание 42iQD
Страница 1: ...42iQD Instruction Manual NOx Analyzer 119013 00 1Sep2021 ...
Страница 2: ......
Страница 32: ...Operation Instrument Display 3 8 42iQD Instruction Manual Thermo Scientific ...
Страница 182: ...Maintenance Pump Rebuilding 5 8 42iQD Instruction Manual Thermo Scientific Figure 5 3 Pump Rebuilding ...
Страница 184: ......
Страница 198: ...Servicing Fan Replacement 7 8 42iQD Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan ...
Страница 212: ...Servicing Power Supply Replacement 7 22 42iQD Instruction Manual Thermo Scientific Figure 7 26 Replacing Power Supply ...
Страница 235: ...Servicing DMC Ozonator Thermo Scientific 42iQD Instruction Manual 7 45 Figure 7 49 Flow Switch Remove ...
Страница 238: ......
Страница 272: ......
Страница 284: ......
Страница 285: ......