Centrifuge GT 2/2R
Introduction and Description | 5
Mains Supply
The following table contains an overview of the electrical connection data for the Centrifuge GT 2/2R. This data is to be taken
into consideration when selecting the mains connection socket.
*
Contact an authorized customer service representative for running the centrifuge at 60 Hz.
Refrigerant
*Contains fluorinated greenhouse gases in a hermetically sealed system.
Keys and display
Easy-to-clean keypad and display surface
Controls
Microprocessor-controlled
Internal memory
The most recent data is saved
Functions
RCF-selection, temperature control, pretemp with cooling equipped devices
Acceleration/braking profiles
9 acceleration and 10 braking curves
Rotor recognition
Automatic
Imbalance recognition
Electronic, contingent on rotor and speed
Lid lock
Automatic lid closing and locking starting from an initial hold position
Table 3. Functions and Features
Component/Function
Description/Features
Table 4. Electrical connection data of the Centrifuge GT 2/2R.
Type
Mains
Voltage
Frequency
Rated
Current
Power
Consumption
Equipment
Fuse
Building
Fuse
Refrigerated
230 V
50/60 Hz
8 A
1 400 W
15 AT
16 AT
Ventilated
230 V
50/60 Hz
6 A
1 010 W
15 AT
16 AT
Table 5. Refrigerant Fisherbrand GT 2R Centrifuge
Order no.
Refrigerant
Quantity
Pressure
GWP
CO
2
e
75004524
R-404A
0.38 kg
31 bar
3 922
2.04 t