Failure messages are shown on the display as codes ER 3 to ER 7. If one of
these codes appears on the display, contact Technical Services
immediately.
To isolate the cause of the failure, perform the following tests.
•
Check to see if the exhaust opening on top of the cabinet is blocked.
•
Ensure that the in-house exhaust air system is activated, if applicable .
•
Close doors and windows in the laboratory to prevent drafts.
•
Turn off devices in the vicinity of the biological safety cabinet that
cause air turbulence, or emit excessive heat.
•
Open flames in the sample chamber may impair airflow conditions.
1300 Series A2
4-5
Thermo Scientific
Section 4
Operation
Error Messages
The observance of work rules ensures a minimum of operational safety
when handling the biological safety cabinet.
Before starting a procedure, take off all jewelry, put on required personal
protective equipment (gloves, goggles, apron), and clean and disinfect the
sample chamber surfaces at regular intervals.
During operation:
•
Place samples only within the defined work area of the work tray.
•
Do not place unnecessary items into the sample chamber.
•
Use only disinfected and cleaned accessories for the work process.
•
Do not cause air turbulence by quick hand, arm or body movements
in the sample chamber or in front of the work opening.
•
Do not place accessories into the sample chamber that cause air
turbulence or emit excessive heat.
•
Do not block air circulation at the ventilation slots of the work tray.
Working
Recommendations
Содержание 1300 Series
Страница 25: ...Thermo Scientific Section 2 Installation 2 8 1300 Series A2 ...
Страница 31: ...Thermo Scientific Section 3 Start Up 3 6 1300 Series A2 ...
Страница 39: ...Thermo Scientific Section 4 Operation 4 8 1300 Series A2 ...
Страница 45: ...Thermo Scientific Section 5 Cleaning and Decontamination 5 6 1300 Series A2 ...
Страница 69: ...Thermo Fisher Scientific 401 Millcreek Road Marietta Ohio 45750 United States www thermofisher com ...
Страница 70: ...1300 系列 A2 II级 A2型 生物安全柜 操作手册 5991355 ...
Страница 91: ...Section 2 安装 维修阀安装 续 维修阀 维修阀插头 垫片 螺母 图2 4 阀门总成和安装 2 6 1300 Series A2 Thermo Scientific 螺纹管 底座 ...
Страница 129: ...仪器类型 部件号 序列号 维修号 地点 操作者标注 技术员 公司 记录 日期 签名 Thermo Scientific 1300 Series A2 9 1 ...